← 返回 JSSC 论文列表
📄 下载 JSSC 原文 PDF
JSSC 2006第12期Other

A 32-Site 4-Channel High-Density Electrode Array for a Cochlear Prosthesis

一种用于人工耳蜗的32位点4通道高密度电极阵列,具有高刺激密度和低功耗特性。
500μA双相电流刺激,8位分辨率,<1%电荷失配,4μs最小脉冲宽度,2.5mW/通道功耗
人工耳蜗高密度电极阵列MEMS技术电流刺激阻抗测量
采用MEMS技术实现高密度电极阵列
集成主动电路实现高精度电流刺激
具备位点阻抗测量和位置感知功能
Abstract
This paper describes a 32-site 4-channel high-den- sity intracochlear electrode array. Combining MEMS-based processing technology with active circuitry, the thin-film device realizes a stimulating site density three times that of commercial systems, delivering 500 A biphasic current levels with 8-bit resolution and less than 1% charge mismatch. The minimum pulsewidth is 4 s, and the per-channel power dissipation is 2.5 mW from 2.5 V. The active circuit chip has a footprint of 2.4 mm 2.9 mm. Ser