A Micro-Power EEG Acquisition SoC With Integrated Feature Extraction Processor f

一款低功耗SoC,集成EEG采集和特征提取,用于癫痫患者持续监测。
1 V, 9 µJ per feature vector, 0.5 Hz feature vector rate
低功耗EEG采集特征提取癫痫监测SoC
采用斩波稳定技术的高输入阻抗仪表放大器
使用静态偏置比较器的低功耗ADC
基于患者数据验证的低功耗特征提取处理器
Abstract
Member , IEEE, Ali Shoeb, Jose Bohorquez , Member , IEEE, Joel Dawson , Member , IEEE, John Guttag, and Anantha P. Chandrakasan , Fellow, IEEE Abstract—This paper presents a low-power SoC that performs EEG acquisition and feature extraction required for contin- uous detection of seizure onset in epilepsy patients. The SoC corresponds to one EEG channel, and, depending on the pa- tient, up to 18 channels may be worn to detect seizures as part of a chronic treatment system. The SoC integrates an