A 104 mW Electrical Impedance Tomography SoC for Portable Real-Time Lung V entil

一款用于便携式实时肺部通气监测的104 mW电阻抗断层成像SoC。
0.18 µm CMOS, 1.8V, 10.4 mW, 20 frames/s
电阻抗断层成像SoC便携式监测肺部通气实时成像
采用T-switch实现60 dB以上的高电极间隔离
I/Q信号生成与解调获取图像实部和虚部
片上快速解调方案降低ADC速度要求以实现低功耗
Abstract
An electrical impedance tomography (EIT) SoC is proposed for the portable real-time lung ventilation monitoring system. The proposed EIT SoC is in tegrated into belt-typefabric system with 32 electrodes and can show the dynamic images of the lung ventilation on the mobile devi ces. To get high fidelity images, a T-switch is adopted for high off-isolation between electrodes more than 60 dB, and I/Q signal generation and demodulation can obtain both real and imaginary part of images. For the real-t