该论文实现了一个用于统一体域网控制器的10.8mW双频段收发器,结合了低功耗体信道通信(BCC)和医疗植入通信服务(MICS),解决了可穿戴和植入式医疗设备在单一无线网络中的通信控制问题。
With the increasing number of portable and implantable personal health care devices, there is a strong demand to control their communication in a single wireless network. Recently, the IEEE 802.15 WBAN task group has discussed the combining of wearable and implantable body sensor networks (BSNs) [1], but no real chip implementation has been reported. In this paper, the implementation of a unified BSN as shown in Fig. 24.9.1 is described.. The unified BSN combines low-power body-channel communication (BCC) [2] and versatile medical implant communication service (MICS) [3] using a network controller located on the human body. This unified BSN has 2 main advantages over the conventional BSNs. First, the MICS band antenna shared with the BCC electrode can be attached directly to human skin to shorten the communication distances among the controller and implanted radios, relaxing their sensitivity and selectivity requirements. Second, due to low path loss of
Namjun Cho, Joonsung Bae, Sunyoung Kim, Hoi-Jun Yoo
KAIST, Daejeon, Korea