← 返回论文列表 📄 下载原文 PDF  ISSCC 2018 · 10.5
ISSCC 2018Session 10 · SENSOR SYSTEMSSensors

A 0.91mW/Element Pitch-Matched Front-End ASIC with Integrated Subarray Beamforming ADC for Miniature 3D Ultrasound Probes

⚡ 本页包含 AI 生成的分析内容,仅供参考

📋 论文概要

该论文提出了一种用于微型3D超声探头的适配间距前端ASIC,集成了子阵列波束成形ADC,有效解决了二维换能器阵列中大量元件与有限电缆数量之间的不匹配问题。

💡 主要创新点

核心指标
0.91mW/Element
重要性
发表年份
ISSCC 2018

🏷 关键词

3D超声探头前端ASIC波束成形ADC低功耗子阵列

📄 原文摘要

Mingliang Tan1, Hendrik J. Vos1,2, Johan G. Bosch2, Martin D. Verweij1,2, Nico de Jong1,2, Michiel A. P. Pertijs1 Delft University of Technology, Delft, The Netherlands Erasmus MC, Rotterdam, The Netherlands 1 2 Data acquisition from 2D transducer arrays is one of the main challenges for the development of emerging miniature 3D ultrasound imaging devices, such as 3D trans-esophageal (TEE) and intra-cardiac echocardiography (ICE) probes (Fig. 10.5.1). The main obstacle lies in the mismatch between the large number of transducer elements (103 to 104) and the limited cable count (<200). Recent advances in transducer-on-CMOS integration have enabled the use of in-probe subarray beamforming based on delay-and-sum (DAS) circuits [1] to reduce the channel count by an order of magnitude. Further reduction calls for in-probe digitization to enable more advanced data processing and compression in the

👥 作者与机构

Chao Chen1, Zhao Chen1, Deep Bera2, Emile Noothout1, Zu-Yao Chang1,

分类:Sensors · 年份:ISSCC 2018