2 imec - Holst Centre, Eindhoven, The Netherlands, Delft University of Technology, Delft, The Netherlands, 3imec, Leuven, Belgium An important drawback of current biopotential monitoring systems is their dependence on gel electrodes, which can dry out, cause skin irritation, and necessitate skilled personnel. These associated drawbacks increase the running costs and significantly hamper their use in consumer healthcare and lifestyle applications. Unfortunately, the use of gel-free, or dry, electrodes increases the electrode-tissue contact impedance, thus exacerbating the effects of interference and cable motion artifacts. A solution is the use of active electrodes, i.e. electrodes in which an amplifier with high input impedance, low noise and good electrode offset rejection is co-integrated (Fig. 17.1.1). Previous active electrodes employed voltage buffers to facilitate the inter-channel gain matching
Jiawei Xu1,2, Refet Firat Yazicioglu3, Pieter Harpe1, Kofi A.A. Makinwa2, Chris Van Hoof3