该论文提出了一款用于心脏再同步治疗的超低功耗心内信号处理芯片,集成了生物阻抗测量和模拟特征提取功能,实现了82dB的动态范围。
of multiple intra-cardiac signals is becoming more and more important for Cardiac Resynchronization Therapy (CRT) and for the analysis of the intra-thoracic fluid status [1, 2]. Robust and accurate Heart-Rate (HR) monitoring at the right/left ventricles and the right atrium is essential for an implantable cardiac pacemaker system (Fig. 16.9.1), and ultra-low power consumption is needed. In addition, an accurate motion sensor and thoracic impedance measurement can acquire valuable additional information in clinical research on the intra-cardiac rhythm analysis. As shown in Fig. 16.9.2, our proposed Analog Signal Processor (ASP) IC consists of 3 power-efficient intra-cardiac signal readout channels. Each channel is equipped with a low-power QRS
Sunyoung Kim1, Long Yan1, Srinjoy Mitra1, Masato Osawa2,
Yasunari Harada2, Kosei Tamiya2, Chris van Hoof1,3, Refet Firat Yazicioglu1 imec, Leuven, Belgium, 2Olympus, Tokyo, Japan, 3KU Leuven, Leuven, Belgium 1 The accurate recognition