该论文提出了一款680nA的全集成植入式心电图采集IC,通过将QRS波段功率参数提取转移到模拟域,显著降低了系统功耗,解决了植入式起搏器对超低功耗和微型化的需求。
Leuven, Belgium 1 2 Ultra-low power consumption and miniature size are by far the most important design requirements for implantable pacemakers. In order to guarantee a long life span of the device, saving power in the sensing IC is a primary concern as cardiac rhythm disorders must be continuously monitored [1]. Shifting the functionality of QRS-band power parameter extraction to the analog domain can reduce system-level power consumption of heartbeat detection significantly through minimizing computational complexity of the DSP [2,3]. In addition, current biomedical ICs still require further improvement of power efficiency as their analog back ends consume significant power [2-4]. For low-power means, the presented analog signal processor (ASP) introduces a power-efficient analog
Long Yan1, Pieter Harpe2, Masato Osawa3, Yasunari Harada3,
Kosei Tamiya3, Chris Van Hoof1,4, Refet Firat Yazicioglu1 imec, Heverlee, Belgium, Eindhoven University of Technology, Eindhoven, The Netherlands, 3 Olympus, Tokyo, Japan, 4 KU Leuven