← 返回论文列表 📄 下载原文 PDF  ISSCC 2014 · 24.8
ISSCC 2014Session 24 · INTEGRATED BIOMEDICAL SYSTEMSMedical & Bio

An Analog-Digital-Hybrid Single-Chip RX Beamformer with Non-Uniform Sampling for 2D-CMUT Ultrasound Imaging to Achieve Wide Dynamic Range of Delay and Small Chip Area

⚡ 本页包含 AI 生成的分析内容,仅供参考

📋 论文概要

该论文提出了一种用于2D-CMUT超声成像的模拟-数字混合单芯片接收波束成形器,采用非均匀采样技术,解决了传统全数字波束成形器功耗高、面积大的问题。

💡 主要创新点

重要性
发表年份
ISSCC 2014

🏷 关键词

超声成像波束成形非均匀采样CMUT模拟-数字混合

📄 原文摘要

Min-Kyun Chae1, Jongkeun Song2, Baehyung Kim2, Seunghun Lee2, Jihoon Bang2, Youngil Kim2, Kyungil Cho2, Byungsub Kim1, Jae-Yoon Sim1, Hong-June Park1 Pohang University of Science and Technology, Pohang, Korea, Samsung Advanced Institute of Technology, Yongin, Korea 1 2 Ultrasound imaging is widely used for medical diagnosis, because it is harmless to the human body and has real-time processing capability. Usually the focusing (beamforming) operation is performed for both TX and RX. The RX focusing is performed by an RX beamformer [1-5], which consists of delay elements and adders. Nowadays, digital beamformers (DBF) are mostly used for conventional ultrasound imaging because of high SNR. Recently, 2D ultrasound transducers have been introduced for 3D imaging. Since the 2D transducer has a huge number of transducer elements (e.g., 9216 for a 72×128 array), it cannot use DBF because of the huge number of required ADCs and wires inside the probe

👥 作者与机构

Ji-Yong Um1, Eun-Woo Song1, Yoon-Jee Kim1, Seong-Eun Cho1,

分类:Medical & Bio · 年份:ISSCC 2014