该论文提出了一种基于有机晶体管的柔性湿传感器片,能够承受2kV ESD,用于生物医学应用(如尿湿检测),并支持无线供电。通过RC振荡器频率变化检测干/湿状态,实现了多传感器复用和无线数据传输。
the reference RC oscillator 1 is 10Hz, RC oscillator 2 (= Sensor 1) is not oscillating, which corresponds to “Dry” (= no urination), and RC oscillator 3 (= Sensor 2) is 5Hz, which corresponds to “Wet” (= urination). Figure 30.3.3(d) shows a successful toggle operation of the multiplexer. Hiroshi Fuketa1,2, Kazuaki Yoshioka1,2, Tomoyuki Yokota1,2, Wakako Yukita1,2, Mari Koizumi1,2, Masaki Sekino1,2, Tsuyoshi Sekitani1,2, Makoto Takamiya1,2, Takao Someya1,2, Takayasu Sakurai1,2 Figure 30.3.4(a) shows a circuit schematic of the RC oscillator with the pseudoCMOS inverters [3]. The RC oscillator converts the resistance between electrodes (RMEA) to the frequency of OSC. Fig. 30.3.4 (b) shows the measured waveform of OSC. Figure 30.3.4(c) shows the measured dependence of the oscillation period on RMEA. In this work, the target range of RMEA is from 2MΩ to 10MΩ because RMEA in cotton immersed in normal saline is several megaohms.
TXdata, RC oscillator 3, and the reference RC oscillator 1. In this measurement,