A 27.8μW Biopotential Amplifier Tolerant to 30Vpp Common-Mode Interference for Two-Electrode ECG Recording in 0.18μm CMOS

⚡ 本页包含 AI 生成的分析内容,仅供参考

📋 论文概要

本文提出了一种用于双电极心电图记录的生物电位放大器,能够容忍高达30Vpp的共模干扰,解决了无参考电极时电源线耦合导致的巨大共模干扰问题。该放大器采用0.18μm CMOS工艺,功耗仅为27.8μW。

💡 主要创新点

核心指标
27.8μW功耗,30Vpp共模干扰容限
工艺节点
0.18μm CMOS
重要性
发表年份
ISSCC 2019

🏷 关键词

生物电位放大器共模干扰双电极ECG

📄 原文摘要

Two-electrode ECG devices have gained popularity in the recent past to enable comfortable and long-term monitoring of cardiovascular health. As a ground or bias electrode is not used in a two-electrode ECG device, common-mode interference (CMI) caused by powerline coupling to the human body can be as large as a few tens of volts. Such a large CMI ruins the ECG recording, and thus the analog front-end of the ECG device must be immune to large CMI. The input common-mode range of a front-end amplifier is limited as the input transistors must operate in the saturation region. Although a CM cancellation path can increase the CM range, it is useful only when the CMI is smaller than the supply voltage [1,3]. For large CMI that exceeds the supply rail, the ESD diode is turned on, and the input is saturated to supply or ground. Unfortunately, current state-of-the-art ECG devices available commercially or in research papers do not