本文提出一种22.6µW的生物电位放大器,采用自适应共模干扰消除技术,解决了双电极心电图记录中因电极阻抗不匹配导致的共模转差模干扰以及电源饱和问题,实现了104dB的总共模抑制比。
Korea Aerospace Research Institute, Daejeon, Korea 1 2 Improving robustness to common-mode interference (CMI) is imperative for reliable two-electrode ECG recording. CMI degrades the signal quality in two ways. First, it appears as a differential-mode signal together with the ECG, because of common-mode to differential-mode (CM-DM) conversion. This is often due to the electrode impedance mismatch [1], as contacts inevitably lose their properties differently during long-term recording. Second, CMI that exceeds the supply voltage saturates the amplifier and distorts the recording when the subject is near power lines [4]. To address both of these issues, the recoding IC must withstand large CMI and also have a large total-CMRR (TCMRR) that includes the effect of electrode mismatch. Unfortunately, prior works provide solutions to only one of them, and thus there is a lack of reliable long-term ECG recording
Nahmil Koo1, Hyojun Kim2, SeongHwan Cho1
KAIST, Daejeon, Korea