⚡ 本页包含 AI 生成的分析内容,仅供参考
本文提出了一种用于干电极的生物阻抗接口芯片,通过电流失配消除和输入增强技术,实现了0.5mΩ/√Hz的低噪声和106dB的高信噪比,面积仅0.45cm²,适用于可穿戴健康监测中的阻抗心动图应用。
Bioimpedance (BioZ) analysis has been recognized as a new paradigm to derive a number of body composition and hemodynamic measures in a non-invasive manner. Measuring the changes in electrical resistance of the thorax during a cardiac cycle, known as impedance cardiography (ICG), is beneficial in detecting early signs of heart failure deterioration [1]. This raises the need for power-efficient wearable BioZ sensors to enable long-term and user-friendly health monitoring, while state-of-the-art designs still suffer from a few drawbacks. First, conventional BioZ interfaces typically rely on gel electrodes (>10cm2) for low-impedance contact between skin and electrodes [1-2]. This not only hampers the long-term recording but also causes user discomfort. However, it is very challenging to perform small-size dry-electrode BioZ sensing at the frequency range of 1kHz to 1MHz, where both the increased electrode-tissue impedance (ETI)
Qinjing Pan, Tianxiang Qu, Biao Tang, Fei Shan, Zhiliang Hong, Jiawei Xu
Fudan University, Shanghai, China