本文提出了一款8通道非接触式心电图记录芯片,采用时分复用技术,能够耐受高达20Vpp的共模干扰,解决了长期穿戴式ECG监测中环境干扰导致的信号质量下降问题。
There has been increasing demand for a low-power wearable device to perform longterm ambulatory monitoring of electrocardiogram (ECG). However, for a wearable device to be a convenient continuous recording system in one’s daily routine, technical issues in ECG become more important. First, long-term monitoring should be robust against massive environmental interference. The interference affects the device in the form of common-mode interference (CMI) caused by coupled disturbances from surrounding powerlines. Common-mode charging/discharging at the input stage has been proposed in 2-electrode systems using switched capacitors [1] and current sources [2] to provide the effective function of a driven-right-leg (DRL) circuit. Second, the device is required to have a high input impedance for high sensitivity. However, there are a few factors that eventually lower the input impedance. A large parasitic capacitance associated with the input path of the instrumentation amplifier (IA)
Kyu-Jin Choi, Jae-Yoon Sim
Pohang University of Science and Technology, Pohang, Korea