本文提出了一款用于导管式高帧率3D超声成像的专用集成电路,集成了65V高压发射器和共享混合波束成形ADC,解决了微型探头中信号质量与连接数量的矛盾。
Zu-Yao Chang1, Chao Chen1, Hendrik Vos1,2, Hans Bosch2, Martin Verweij1,2, Nico de Jong1,2, Michiel Pertijs1 Delft University of Technology, Delft, The Netherlands Erasmus MC, Rotterdam, The Netherlands 1 2 Intra-cardiac echography (ICE) probes (Fig. 32.2.1) are widely used in electrophysiology for their good procedure guidance and relatively safe application. ASICs are increasingly employed in these miniature probes to enhance signal quality and reduce the number of connections needed in mm-diameter catheters [1-5]. 3D visualization in real-time is additionally enabled by 2D transducer arrays with, for each transducer element, a highvoltage (HV) transmit (TX) part, to generate acoustic pulses of sufficient pressure, and a receive (RX) path, to process the resulting echoes. To achieve the required reduction in RX channels, micro-beamforming (μBF), which merges the signals from a subarray using a delay-and-sum operation, has been shown to be an effective solution [3,4].
Yannick Hopf1, Boudewine Ossenkoppele1, Mehdi Soozande2, Emile Noothout1,