← 返回论文列表 📄 下载原文 PDF  ISSCC 2025 · 15.7
ISSCC 2025Session 15 · NEURAL INTERFACES AND EDGE INTELLIGENCE FOR MEDICAL DEVICESAI / ML

A 4.6µW 3.3-NEF Biopotential Amplifier with 133VPP Common-Mode Interference Tolerance and 102dB Total Common-Mode Rejection Ratio for Two-Electrode Recording System

⚡ 本页包含 AI 生成的分析内容,仅供参考

📋 论文概要

本文提出了一种用于双电极生物电位记录的低功耗放大器,功耗仅4.6µW,噪声效率因子为3.3。该放大器能够承受高达133VPP的共模干扰并实现102dB的总共模抑制,解决了传统双电极记录中因大共模干扰导致的饱和或共模转差模问题。

💡 主要创新点

核心指标
4.6µW功耗, 3.3 NEF, 133VPP共模干扰容限, 102dB总CMRR
重要性
发表年份
ISSCC 2025

🏷 关键词

生物电位放大器共模干扰容限低功耗双电极记录噪声效率因子

📄 原文摘要

crucial in delivering vital information for medical diagnostics and research applications. Recently, the demand for biopotential recording using two electrodes has grown thanks to its better user experience and lower cost than counterparts with three electrodes [1,2]. However, a two-electrode recording IC suffers from a large common-mode interference (CMI) over 100VPP [3], potentially saturating an analog front-end (AFE) or resulting in large CM to differential-mode (DM) conversion. These challenges necessitate biopotential AFEs to possess a large CMI tolerance as well as a high total common-mode rejection ratio (T-CMRR) while providing excellent noise efficiency with low power consumption. Figure 15.7.1(a) depicts a simplified electrical

👥 作者与机构

Yongjae Park1, Yeong-Jin Mo2, Jeong-Hoon Kim3, Gert Cauwenberghs3, Seong-Jin Kim2

Ulsan National Institute of Science and Technology, Ulsan, Korea Sogang University, Seoul, Korea 3 University of California, San Diego, CA 1 2 Physiological data, such as EEG and ECG, are

分类:AI / ML · 年份:ISSCC 2025