A 30MHz Wideband 92.7dB SNR 99.6% Accuracy Bioimpedance Spectroscopy IC Using Time-to-Digital Demodulation with Co-Prime Delay Locked Sampling

⚡ 本页包含 AI 生成的分析内容,仅供参考

📋 论文概要

该论文提出了一款宽带生物阻抗频谱测量IC,采用时间数字解调与互质采样技术,将工作频率从传统的10MHz扩展到30MHz,实现了92.7dB SNR和99.6%的精度,解决了现有生物阻抗IC带宽不足的问题。

💡 主要创新点

核心指标
30MHz带宽, 92.7dB SNR, 99.6%精度
重要性
发表年份
ISSCC 2025

🏷 关键词

生物阻抗频谱时间数字解调互质采样宽带高精度

📄 原文摘要

The permittivity and conductivity of bioimpedance in tissues are governed by α, β, and γ dispersions, with biological features detectable in the tens of MHz range. Wideband spectroscopy is required in many applications including body fluid volume assessment [1], cancer diagnosis [2], anatomical characterization [3], and plant content estimation [4]. However, state-of-the-art bioimpedance ICs are limited to frequencies up to 10MHz, which hampers biological research.%For example, estimating starch content in cassavas requires wideband spectroscopy with frequencies up to 30MHz [4]. In a wideband system using analog I-Q demodulation, accuracy is limited by the large distortion from its multiplication process; a phase error of 4.32° at 10MHz was reported in [2]. In a wideband system using digital I-Q demodulation, a phase error of 0.15% at 10MHz was reported in [5] but at the expense of a large power consumption (12.2mW for the amplifier and ADC alone) and a

👥 作者与机构

Jiayang Li, Dai Jiang, Yu Wu, Andreas Demosthenous

University College London, London, United Kingdom

分类:Medical & Bio · 年份:ISSCC 2025